Home > Flammability Test Equipment > UL910 Steiner Tunnel Furnace chamber
UL910 Steiner Tunnel Furnace chamber
UL910 Steiner Tunnel Furnace chamber

UL910 Steiner Tunnel Furnace chamber

Product Model : SL-UL910
UL Standard for Safety Test for Flame-Propagation and Smoke-Density Values for Electrical and Optical-Fiber Cables Used in Spaces Transporting Environmental Air
Detailed description

UL910 Steiner Tunnel Furnace chamber


Design referenced to the below standards

UL 910 : UL Standard for Safety Test for Flame-Propagation and Smoke-Density Values for Electrical 

and Optical-Fiber Cables Used in Spaces Transporting Environmental Air

Steiner Tunnel Furnace Chamber Design Proposal

2.1   Summary 

This technical solution is based on the integration of instrumentation and electricity, and uses OMRON's advanced multi-function control system to 

combine ignition sequence, combustion safety, electrical interlocking, automatic temperature control,manual adjustment, monitoring alarm, and 

data acquisition / communication into one. The furnace temperature and pressure control system uses the control system composed of 

OMRON controller + Advantech + Visual Basic software for automatic control and online monitoring. 

The upper computer Advantech operation station operates and monitors the necessary process parameters of the furnace. 

It has data storage, inspection and printing. , And other functions. At the same time, according to the requirements of UL910 / NFPA262 standards, 

a dedicated combustion test room can be set up for combustion test, This combustion test room effectively isolates the Steiner horizontal tunnel furnace,

Install the smoke density measuring end in a dark room,Avoid interference from external light;Burn test area, adopt independent design method,according to the standard requirement, should provide free airflow. Therefore, during the entire test, keep the room under controlled air pressure of 

0-12 Pa (0-0.05 inch water volume) higher than the surrounding air pressure.,The temperature is maintained at 18.3 ° C-26.7 ° C (65 ° F-80 ° F) and 

a relative humidity of 45-60%.

2.1.1 Design conditions

 Furnace type: Steiner horizontal tunnel furnace

 Furnace main dimensions: Furnace size 7620mm * 451mm * 305mm

 Number of furnace covers: Steiner horizontal tunnel furnace: 1 furnace cover

 Furnace working temperature: up to 600 ℃ (flue gas temperature)

 Fuel: Methane with a purity of at least 95%

 Fuel calorific value: 3500btu / lb

 Fuel pressure: 0.4-0.5MPa

 Burner model: 3/4 inch U-shaped double burner

 Exhaust temperature of flue pipe: <250 ° C, usually the flue gas temperature is controlled within 200 ° C.

 Test method of the tested object: Horizontal component furnace-furnace ceiling hoisting

   Design conditions 15KVA, 380V/220V, 3 phases. Notes: Voltage can be customized.

2.1.2 Structural parameters

2.1.3 Furnace structure: refractory brick + SUS304 stainless steel plate

2.1.4 Furnace bottom structure: refractory brick 229mm x 114.5mm x 64mm 

2.1.5 Type and quantity of burner: 1 3 / 4-inch U-shaped double burner.

2.1.6 Smoke exhaust method: mechanical smoke exhaust + cold air mixing on the back wall

2.1.7 Furnace door opening method: Horizontal component furnace-hoist ceiling hoisting and translation (customer built)

2.1.8 Furnace diagram

2.1.4  Purpose:          

For wire and cable UL910 flammability testing

2.1.5 Design principles

       Adopt the principles of advanced technology, reliability, safety, and economic rationality 

2.2   Furnace structure

2.2.1 Furnace Shell

The furnace steel structure is composed of square tube / rectangular tube end, side upright and furnace body steel plate. 

After being welded with high strength, it forms a solid whole, which can be used without deformation for a long time.

 Furnace body steel plate:SUS304,δ=3mm

 Furnace frame:Q235-A ,Square tube / rectangular tube

 Furnace transverse bars:Q235-A ,Square tube/ rectangular tube

 Window:Quartz glass and tempered glass double layer combination,δ=3mm,70mm±6mm×280mm±38mm

2.2.2 Furnace Resistance materials

The test furnace masonry is consisted of refractory bricks, At the same time, in order to provide the air turbulence required during 

the combustion process,Obtained by placing six 229mm long x 114.5mm wide x 64mm thick heat-resistant refractory bricks

(Long vertical line of the wall and 114.5mm long parallel line). According to the measured burner centerline to the refractory 

brick centerline,Refractory bricks near the window (without obstructing the windows) 1.98m±152mm、3.96m±152mm and 5.79±152mm,

the distance of another side is 1.37m±152mm、2.90m±152mm and 4.88m±152mm.

 The highest Temperature resistance:    1427℃   (2600℉)

 Bulk Density:    0.77±0.046g/cm3

 Thermal conductivity at average temperature:

 260℃(500℉)     0.23W/m·℃                   

 538℃(1000℉)    0.27 W/m·℃                 

 815℃(1500℉)    0.32 W/m·℃               

 1093℃(1500℉)    0.37 W/m·℃       

2.2.3 Furnace door and Pressing mechanism

Operation Furnace door above the furnace body, as a sealing device for the furnace body. Composed of metal and inorganic insulators, Insulators composed of inorganic insulators, 51mm ± 6mm thick inorganic insulator material

Horizontal furnace door is welded by section steel, use self-weight vertical lifting method to compact, to observe the condition of the furnace, observation windows installed on both sides of the furnace wall, to get an effective seal, an effective water seal acts as a seal between the furnace door and the furnace body, Using tap water as the circulating water source can not only provide a seal for the quality inspection of the furnace body and the furnace door, but also take away the heat during the combustion test and effectively protect the furnace body.

The lab should provide the crane to lift the lid.   

 Max. Effective use temperature up to 1050℃;

 Density: 335±5kg/m3;

 Thermal Conductivity: 0.085W/mK@400 ℃

 Dimension 7620±50mm*451±5mm*305±5mm

2.2.4 Inlet chamber and inlet baffle

The furnace steel structure is composed of square tube / rectangular tube end, side upright and furnace body steel plate. 

After being welded with high strength, it forms a solid whole, which can be used without deformation for a long time. 

The air intake baffle is pneumatically controlled and can be opened and closed automatically. Intake chamber 

This element shall have a rectangular opening of 298.5mm ± 6mm × 464mm ± 6mm to allow air to pass through 

the nearest baffle into the combustion test chamber.

 Furnace steel plate: SUS304, δ = 3mm

 Furnace frame: Q235-A, square tube / rectangular tube

 Furnace transverse ribs: Q235-A, square tube / rectangular tube

 


2.2.4 Smoke exhaust system and furnace pressure control system

The exhaust of the furnace body adopts the form of mechanical exhaust to ensure that the pressure and temperature in 

the furnace and the exhaust gas meet the standard requirements. It includes transition section, smoke exhaust pipeline, 

automatic butterfly valve, and differential pressure control system. 

Transition section: a rectangular section stainless steel element with a length of 

902mm ± 6mm × 686mm ± 6mm in width × 438mm ± 6mm in height,and a 457mm±6mm 

It consists of a long rectangular elliptical transitional section, and the rectangular elliptical transitional section is connected to 

an exhaust pipe with an inner diameter (I.D.) of 406mm ± 3m. 

The exterior of the transitional section is insulated with a 51mm ceramic fiber cover, with a density of 130kg / m3. 

The steel plate is SUS304, δ = 1.5mm. 

Exhaust pipe: 406mm ± 3mm I.D. exhaust pipe, extending from the exhaust end of the transition section from 4.88m to 5.49m to 

the center line of the smoke measurement system, so as to provide a completely mixed exhaust gas flow. 

The exhaust pipe opening shall be insulated with a high temperature inorganic material of at least 51mm, from the beginning of 

the exhaust transmission part to the smoke detection system. 

The steel plate is SUS304, δ = 1.5mm. 

Differential pressure control system: 

The detector should consist of a stainless steel column with a column length rated twice the outer diameter of the column, 

the length of the draft gauge tap is 25±12mm. 25±12mm, and a central solid partition. 

The bidirectional probe is connected to a pressure sensor, which can effectively read the pressure value in the furnace.

 



Exhaust damper:406mm I.D. 

The single-piece pipe flow control damper of the pipe is installed at 1.68m ± 0.15m below the exhaust pipe of 

the smoke measurement system, and the center line is to the center line.

The relative positions of exhaust transition components exhaust ducts, smoke measurement systems, 

and exhaust duct dampers are shown in the figure.

In order to maintain control of the air flow during the entire test process, 

the exhaust pipe damper should be controlled by a closed-loop feedback system that forms effective communication with 

the differential pressure control system.



2.2.5  Effect drawing:Steiner Tunnel Furnace Chamber